VIGAZ BOX100/120-USB analyzer

What it is for and who uses it

The BOX100/120-USB is an analyzer for measuring modified atmospheres in food packaging, in BOX100 (O₂) or BOX120 (O₂ and CO₂) version, with a traceability function. According to the VIGAZ data sheet, it stores measurements in its internal memory and the data are recorded in a file that can be used directly in Excel without any external software. The operator can program different products to retrieve one file per product and per day. The connected USB key retrieves the data, empties the memory and avoids any cabling.

Specifications

According to the VIGAZ technical data sheet:

BOX100/120-USB specifications
ParameterValue
Measured gasesO₂ (BOX100) · O₂ and CO₂ (BOX120)
O₂ sensorElectrochemical · range 0-100 % · accuracy ±1 % relative · resolution 0.1 % · response (T90) 10 s · service life 4 years in air
CO₂ sensor (BOX120)Infrared · range 0-100 % · accuracy ±2% · resolution 0.1 % · response (T90) 9 s · service life unlimited
Operating temperature0 to +40 °C
ProtectionIP65 (closed)
Display8-line digital
ConnectivityUSB / 500 records
Data loggingMeasurements are stored in the internal memory and recorded in a file that can be used directly in Excel without external software · different products can be programmed to obtain one file per product and per day · the connected USB key retrieves the data, empties the memory and avoids any cabling
Power supply230 V - 50/60 Hz
Rechargeable battery option12 V battery · charger 100-240 V - 50/60 Hz · battery life: BOX100 > 48 h, BOX120 > 12 h
Dimensions24 x 19 x 11 cm
Weight2.5 kg
FunctionsAutomatic O₂ calibration in air at start-up (20.95 %) · calibration menu protected by a 4-digit security code · product name recording · values locked after each measurement · values stored in memory · blocked gas circuit detection (a light comes on if the filter or needle is blocked) · pump with programmable delay from 0 to 20 s; automatic mode: first 5 s suction and, if the measurement has not stabilised, a second 5 s suction · low sample withdrawal volume · long-life O₂ sensor
CalibrationWith standard gases: O₂ zero / O₂ calibration
Supplied withSpiral sampling kit · foams, filters and needles · calibration certificate and instructions for use

Scroll horizontally to see the whole table.

Data according to the manufacturer's (VIGAZ) technical data sheet.

How it is used, according to the data sheet: insert the needle through a septum and press "P"; the pump draws the sample for the programmed time, an hourglass shows the progress of the measurement and, once the reading has stabilised, it is locked on the display and the values are stored in memory. There is no need to draw in air between one measurement and the next.

Position in the range

What sets it apart, according to the VIGAZ data sheets: it is the only model in the range that records measurements in memory and exports them via a USB key. It has an 8-line digital display, is powered at 230 V - 50/60 Hz and offers a rechargeable 12 V battery option. Its O₂ resolution is 0.1 %, compared with 0.01 % for the BOX100 II and BOX120 II.

Documentation and contact

Request a quotation or the BOX100/120-USB technical data sheet.

Enquire about the BOX100/120-USB

Request a quotation or the BOX100/120-USB data sheet.

VIGAZ use and documentation

According to VIGAZ, BOX100/120-USB adds a traceability function with USB export. Explore the analyzer range and the MAP oxygen analyzer guide.