BOX120-b II: battery-powered O₂/CO₂ analyzer (VIGAZ)
What it is for and who uses it
The BOX120-b II is a dual analyzer measuring O₂ and CO₂ in modified atmosphere packaging and runs on a 12 V battery. According to the VIGAZ data sheet, it is designed to be installed directly at the end of the production line or in the laboratory and to carry out a large number of measurements. A measurement is started at the press of a button and stops when the sensors have finished reacting; the values are then frozen, so each measurement is almost automatic.
Specifications
According to the VIGAZ technical data sheet:
| Parameter | Value |
|---|---|
| Measured gases | O₂ and CO₂ |
| O₂ sensor | Electrochemical · range 0-100 % · accuracy ±1 % relative · resolution 0.01 % · response (T90) 10 s · service life 4 years in air |
| CO₂ sensor | Infrared · range 0-100 % · accuracy ±2% · resolution 0.1 % · response (T90) 9 s · service life unlimited |
| Operating temperature | 0 to +40 °C |
| Protection | IP65 (closed) |
| Display | 4-line digital |
| Power supply | 12 V battery |
| Charger | 100-240 V - 50/60 Hz |
| Battery life | > 12 h |
| Dimensions | 24 x 19 x 11 cm |
| Weight | 2.6 kg |
| Functions | Automatic O₂ calibration in air at start-up (20.95 %) · calibration menu protected by a 4-digit security code · values locked after each measurement · blocked gas circuit detection (a light comes on if the filter or needle is blocked) · pump with programmable delay from 0 to 20 s; automatic mode: first 5 s suction and, if the measurement has not stabilised, a second 5 s suction · low sample withdrawal volume · long-life O₂ sensor |
| Calibration | With standard gases: O₂ zero / O₂ calibration |
| Supplied with | Spiral sampling kit · foams, filters and needles · calibration certificate and instructions for use |
Scroll horizontally to see the whole table.
Data according to the manufacturer's (VIGAZ) technical data sheet.
How it is used, according to the data sheet: insert the needle through a septum and press "P"; the pump draws the sample for the programmed time, an hourglass shows the progress of the measurement and, once the reading has stabilised, it is locked on the display. There is no need to draw in air between one measurement and the next.
Position in the range
What sets it apart, according to the VIGAZ data sheets: it measures O₂ and CO₂ like the BOX120 II and has the same dimensions (24 x 19 x 11 cm), but it runs on a 12 V battery (charger 100-240 V - 50/60 Hz) with more than 12 hours of battery life, whereas the BOX120 II is powered at 230 V - 50/60 Hz. It weighs 2.6 kg, compared with 2.2 kg for the BOX120 II. To measure O₂ only on battery power, the range includes the BOX100-b II.
Documentation and contact
Request a quotation or the BOX120-b II technical data sheet.
Enquire about the BOX120-b II
Request a quotation or the BOX120-b II data sheet.
VIGAZ use and documentation
According to VIGAZ, BOX120-b II runs on a 12 V battery and is intended for production or laboratory use. Explore the analyzer range and the MAP oxygen analyzer guide.